TRANSMITTER,TEMPERATURE,ELECTRIC. Key specifications include CASE TYPE: HINGED LID CARRYING CASE W/HANDLE, UNIT TYPE: PROTECTIVE TUBE, OVERALL LENGTH: 470.0 MILLIMETERS NOMINAL MM, OUTSIDE DIAMETER: 7.5 MILLIMETERS NOMINAL MM, PRECIOUS MATERIAL: GOLD AND PLATINUM AND SILVER, OPERATING TEMP RANGE: -200.0 TO 500.0 CELSIUS DEGC. This product is currently in active production with full OEM support.
This part is cataloged in the Federal Logistics Information System (FLIS) under NSN 6685-00-526-8241, indicating validated use in U.S. government and defense logistics supply chains. Available through authorized enterprise resellers and TPM (Third Party Maintenance) channels.
Verify compatibility with your specific system configuration before ordering.
| CASE TYPE | HINGED LID CARRYING CASE W/HANDLE |
| UNIT TYPE | PROTECTIVE TUBE |
| OVERALL LENGTH | 470.0 MILLIMETERS NOMINAL MM |
| OUTSIDE DIAMETER | 7.5 MILLIMETERS NOMINAL MM |
| PRECIOUS MATERIAL | GOLD AND PLATINUM AND SILVER |
| OPERATING TEMP RANGE | -200.0 TO 500.0 CELSIUS DEGC |
| FUNCTIONAL DESCRIPTION | USED TO MEASURE THE TEMPERATURE OF TRIPLE POINT AND FREEZING POINT CELLS USED IN THE CALIBRATION OF THERMOCOUPLES AND OTHER TYPES OF THERMOMETERS |
| TERMINAL TYPE AND QUANTITY | 4 LUG |
| REFERENCE DATA AND LITERATURE | MFG COMMERCIAL CATALOG |
| PRECIOUS MATERIAL AND LOCATION | TERMINALS GOLD AND RESISTOR AND INTERNAL LEADS TO HANDLE PLATINUM AND SILVER PLATED COPPER LEAD WIRES SILVER |
| ACCOMMODATED CONTACT TYPE AND QUANTITY | 4 LUG |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | STANDARD PLATINUM RESISTANCE THERMOMETER; MEYERS THERMOMETER |
| NSN | 6685-00-526-8241 |
| REFERENCE DATA AND LITERATURE | MFG COMMERCIAL CATALOG |
| OPERATING TEMP RANGE | -200.0 TO 500.0 CELSIUS DEGC |
| SPECIAL FEATURES | LENGTH 470MM EXCLUDING HANDLE; SILVER PLATED COPPER LEADS 2.4M LONG; IMMERSION DEPTH 7 TO 14 INCHES; SESNSING LENGTH 33MM; TIME CONSTANT 6 SECONDS; RESISTANCE AT 0 DEG C IS 25.5 OHMS' SHEATH MATERIAL IS QUARTZ, FILLED WITH DRY AIR; ACCUARCY IS WITHIN 1 MICRO KELVIN MM |
| FUNCTIONAL DESCRIPTION | USED TO MEASURE THE TEMPERATURE OF TRIPLE POINT AND FREEZING POINT CELLS USED IN THE CALIBRATION OF THERMOCOUPLES AND OTHER TYPES OF THERMOMETERS |
| CASE TYPE | HINGED LID CARRYING CASE W/HANDLE |
| PRECIOUS MATERIAL AND LOCATION | TERMINALS GOLD AND RESISTOR AND INTERNAL LEADS TO HANDLE PLATINUM AND SILVER PLATED COPPER LEAD WIRES SILVER |
| PRECIOUS MATERIAL | GOLD AND PLATINUM AND SILVER |
| UNIT TYPE | PROTECTIVE TUBE |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | STANDARD PLATINUM RESISTANCE THERMOMETER; MEYERS THERMOMETER |
| TERMINAL TYPE AND QUANTITY | 4 LUG |